Molecular Devices AquaMax 2000 Microplate Washer

PDFPrintEmail
Two-fluid inlets; 96- and 384-plate wash head, 384-cell wash head included.
2015_06_25_MolecularDevices_AquaMax2000_02
2015_06_25_MolecularDevices_AquaMax2000_03
Description

System includes the following heads:

  • 96- and 384-well Plate Wash Heads for biochemical assays for thorough washing of ELISAs
  • 384-well Cell Wash Head for cell-based assays with angled dispensing pins for gentle washing of cell-based assays

Performance Specifications

Precision (96-well):  < 4% CV @ 300 µL
Precision (384-well):  ≤ 3% CV @ 100 µL
Washing times: 96 wash head, 300 µL/well
Single cycle:  11 sec.
Three cycles:  26 sec.
Washing times 384 wash head, 100 µL/well
Single cycle:  12 sec.
Three cycles:  28 sec.
Dispense volume: 10-400 µL per well
FastWash times (optimized for speed): 96 wash head, 300 µL/well, three cycles: 14 sec.; 384 Wash Head, 100 µL/well, three cycles 16 sec.
Process time with integrated, optional StakMax® Microplate Handling System (Fast Wash, 96 Wash Head, 300 µL/well, three cycles): < 13 min. for 20 plates
Wash volume: 100-3000 µL
Residual volume, 96: ≤ 5 µL
Residual volume, 384: ≤ 2 µL

General Specifications

Dimensions (in.): 8.3 (H) x 18.3 (W) x 20.3 (D)
Dimensions (cm): 21 (H) x 46.6 (W) x 51.7 (D)
Weight (instrument): 40.8 lbs. (18.5 kg)
Power supply: Auto-switching
Wash head: PET (polyethylene terephthalate), 304 stainless steel
Liquid contact materials: Nylon, Polypropylene, Teflon, Polyethylene, Stainless Steel, and Silicone
Compatible microplates: SBS-conforming 96- and 384-well; flat, U-, or V-bottom plates
Outlets: Waste to drain or vented container
Interface connections; USB and proprietary StakMax connect
Shake options: 3 speeds; Up to 600 sec. max.

 